■ Flexibility facilitating conductor clamping even in narrow space
■ AC current measurable max. 3000A
■ True RMS■ Data hold, Max/Min value hold
■ BacklightSampling rate : approx. 2 times / sec.
تشاهد 1–24 من94 النتائج
AC Current | 10AAC |
AC Voltage | 1000V AC |
Capacitance | 100µF |
DC Current Measurement | 10A DC |
DC Voltage | 1000V DC |
Resistance | >40MΩ |
AC Current Resolution | 0.1µA ac |
AC Voltage Resolution | 1mV ac |
Battery Type | AA |
Best AC Current Accuracy | ±1.8% + 5 Digits |
Best AC Voltage Accuracy | ±1.2% + 7 Digits |
Best Capacitance Measurement Accuracy | ±5% + 10 Digits |
Best DC Current Accuracy | ±1.4% + 3 Digits |
Best DC Voltage Accuracy | ±0.5% + 2 Digits |
Best Resistance Measurement Accuracy | ±1.2% + 5 Digits |
Capacitance Measurement Resolution | 10pF |
DC Current Measurement Resolution | 0.1µA dc |
DC Voltage Measurement Resolution | 0.1mV dc |
Dimension | 44 x 82 x 166mm |
Display Type | LCD |
Functions Measured | AC Current, AC Voltage, Capacitance, Continuity, DC Current, DC Voltage, Diode, Frequency, Resistance |
Height | 166mm |
Length | 44mm |
Maximum Frequency | 400Hz |
Model Number p | CD771 |
Resistance Measurement Resolution | 100mΩ |
Safety Category | CAT II 1000 V, CAT III 600 V |
Safety Category Level | CAT II, CAT III |
Safety Category Voltage | 600 V, 1000 V |
True RMS | Yes |
Weight | 360g |
Width | 82mm |
DC voltage (V) | 400mV/4V/40V/400V/1000V | ±(0.5%+5) |
AC voltage (V) | 4V/40V/400V/750V | ±(0.8%+3) |
DC current (A) | 40uA/400uA/4mA/40mA/400mA/10A | ±(0.8%+3) |
AC current (A) | 40mA/400mA/10A | ±(1%+2) |
Resistance (Ω) | 400Ω/4000Ω/40kΩ/400kΩ/4MΩ/200MΩ | ±(0.8%+2) |
Capacitance (F) | 4nF/40nf/400nF/4uF/40uF/400uF/4mF/10mF | ±(4%+5) |
Features | ||
Display count | 4000 | |
Transistor hFE | √ | |
Diode | √ | |
Frequency response | 40Hz~400Hz | √ |
Range selection | Manual | |
Continuity test | √ | |
Data hold | √ | |
Backlight | √ | |
Low battery indication | √ | |
Auto power off | √ | |
General Characteristics | ||
Power | 1.5V battery (R03)×2 | |
Display | 56.5mm x 36mm | |
Product color | Red and grey | |
Product net weight | 330g | |
Product size | 175mm x 83mm x 53mm | |
Standard accessories | Test leads |
AC Current 40A/600A ±(2.5%+5)
DC Current 40A/600A ±(2.5%+5)
AC Voltage 4V/40V/400V/600V ±(1%+5)
DC Voltage 400mV/4V/40V/400V/600V ±(0.8%+1)
Resistance 400Ohm/4KOhm/40KOhm/400KOhm/4MOhm/40MOhm ±(1%+2)
Capacitance4nF/40nF/400nF/4µF/40µF/100µF±(4%+3)
Frequency
10Hz ~ 10MHz ±(0.5%+3)
10MHz Reference only
Temperature -40 °C ~ 1000°C ±(2.5%+3)
Features
Display Count 3999
Auto Range
Jaw Capacity 28mm
Diode
Auto Power Off
Continuity Buzzer
Low Battery Indication
Data Hold
MAX Mode
Setup
Input Impedance for DCV Around 10MOhm
General Characteristic
Power 9V Battery(6F22)
LCD Size 35.6X 18mm
Product Color Red and Grey
Product Net Weight 260g
Product Size 208mm X 76mm X30 mm
Standard Accessories Battery, Test Lead,Carrying Bag,Point Contact Temperature Probe, Gift Box, English Manual
C millivolts | Range/resolution | 600.0 mV / 0.1 mV |
Accuracy | ± ([% of reading] + [counts]): 0.5% + 2 | |
DC volts | Range/resolution | 6.000 V / 0.001 V 60.00 V / 0.01 V 600.00 V / 0.1 V |
Accuracy | ± ([% of reading] + [counts]): 0.5% + 2 | |
AC millivolts1 True RMS | Range/resolution | 600.0 mV / 0.1 mV |
Accuracy | 1.0% + 3 (DC, 45 Hz to 500 Hz) 2.0% + 3 (500 Hz to 1 kHz) |
|
AC volts1 True RMS | Range/resolution | 6.000 V / 0.001 V 60.00 V / 0.01 V 600.0 V / 0.1 V |
Accuracy | 1.0% + 3 (DC, 45 Hz to 500 Hz) 2.0% + 3 (500 Hz to 1 kHz) |
|
Continuity | Range/resolution | 600 Ω / 1 Ω |
Accuracy | Beeper on 250 Ω detects opens or shorts of 500 μs or longer. |
|
Ohms | Range/resolution | 600.0 Ω / 0.1 Ω 6.000 kΩ / 0.001 kΩ 60.00 kΩ / 0.01 kΩ 600.0 kΩ / 0.1 kΩ 6.000 MΩ / 0.001 MΩ |
Accuracy | 0.9% + 1 | |
Range/resolution | 40.00 MΩ / 0.01 MΩ | |
Accuracy | 5% + 2 | |
Diode test | Range/resolution | 2.00 V / 0.001 V |
Accuracy | 0.9% + 2 | |
Capacitance | Range/resolution | 1000 nF / 1 nF 10.00 μF / 0.01 μF 100.0 μF / 0.1 μF 9999 μF / 1 μF 100 μF to 1000 μF |
Accuracy | 1.9% + 2 | |
Range/resolution | > 1000 μF | |
Accuracy | 5% + 20% | |
Lo-Z capacitance | Range | 1 nF to 500 μF |
Accuracy | 10% + 2 typical | |
AC amps True RMS (45 Hz to 500Hz) |
Range/resolution | 6.000 A / 0.001 A 10.00 A / 0.01 A |
Accuracy | 1.5% + 3 20 A overload for 30-seconds max |
|
DC amps | Range/resolution | 6.000 A / 0.001 A 10.00 A / 0.01 A |
Accuracy | 1.0% + 3 20 A overload for 30-seconds max |
|
Hz (V or A input)2 | Range/resolution | 99.99 Hz / 0.01 Hz 999.9 Hz / 0.01 Hz 9.999 kHz / 0.001 kHz 50.00 kHz / 0.01 kHz |
Accuracy | 0.1% + 2 |
AC A | 0 – 600.0/1000A ±1.5%rdg±5dgt (50/60Hz) ±3.0%rdg±5dgt (40 – 400Hz) |
---|---|
DC A | 0 – 600.0/1000A ±1.5%rdg±5dgt |
AC V | 6/60/600V Auto Ranging ±1.3%rdg±4dgt (50/60Hz) ±3.0%rdg±5dgt (40 – 400Hz) |
DC V | 600m/6/60/600V Auto Ranging ±1.0%rdg±3dgt |
Ω | 600/6k/60k/600k/6M/60MΩ (Auto Ranging) ±1%rdg±5dgt (600 – 6M) / ±5%rdg±8dgt (60M) |
Continuity buzzer | Buzzer Sounds at 100Ω |
Frequency/DUTY | 10/100/1k/10kHz (Auto Ranging) / 0.1 – 99.9% |
Conductor size | Ø40mm |
Voltage LED Steps (AC/D):
– 12 V (between 0 to 12 V LED will illuminate)
– 24 V (between 19 V +/- 2 V LED will illuminate)
– 50 V (between 48 V +/- 6 V LED will illuminate)
– 120 V (between 110 V +/- 20 V LED will illuminate)
– 230 V (between 200 V +/- 20 V LED will illuminate)
– 400 V (between 340 V +/- 30 V LED will illuminate)
– 690 V (between 520 V +/- 30 V LED will illuminate)
Continuity test: Range 0 to 100 kOhm. The continuity LED will illuminate and buzzer will sound
Phase Rotation Test: 50Hz to 60Hz
Range Selection: Automatic
Polarity Detection: Positive & Negative
Low Battery Display: Short circuit two leads, LED and beeper do not on that means the battery is low
Torch: Yes
Single Lead (L2) Voltage Detection: It is intended only as a quick check. 100V to 690V, 50Hz to 400Hz
Short Circuit Detection: Yes
GENERAL CHARACTERISTICS:
Power: 2x 1.5 V battery (AAA)
LCD display (AC/DC): 24V to 690V (+/- 3%+5)
Product Net Weight: 210 g
Product size: 275 x 51 x 30 mm
AC A | 400A/2000A (0 – 1000A) ±1.5%rdg±2dgt [50/60Hz] ±3%rdg±4dgt [40 – 500Hz] ±5%rdg±4dgt [500Hz – 1kHz] 2000A (1001 – 2000A) ±3%rdg±2dgt [50/60Hz] |
---|---|
DC A | 400/2000A ±1.5%rdg±2dgt |
AC V | 400/750V ±1.5%rdg±2dgt [50/60Hz] ±1.5%rdg±4dgt [40Hz – 1kHz] |
DC V | 400/1000V ±1%rdg±2dgt |
Ω | 400/4000Ω ±1.5%rdg±2dgt |
Continuity buzzer | buzzer sounds below 50±35Ω |
Conductor size | Ø55mm max. |
Frequency response | 40Hz – 1kHz |
Output | Recorder : DC400mV against AC/DC400A DC200mV against AC/DC2000A |
Safety Standard | IEC 61010-1 CAT IV 600V , CAT III 1000V IEC 61010-031 IEC 61010-2-032 |
Power source | R6 (AA) (1.5V) × 2 |
Dimensions | 250(L) × 105(W) × 49(D)mm |
Weight | 530g approx. |
Included Accessories | 7107A (Test leads) 9094 (Carrying Case) R6 (AA) × 2 Instruction Manual |
DC Voltage: 400mV / 4V / 40V / 400V / 600V
– best accuracy: +/- (0.8%+1)
AC Voltage: 4V / 40V / 400V / 600V
– best accuracy: +/- (1%+5)
DC Current: 40A / 400A
– best accuracy: +/- (2%+3)
AC Current: 40A / 400A
– best accuracy: +/- (2.5%+5)
Resistance: 400 Ohm / 4 kOhm / 40 kOhm / 400 kOhm / 4 MOhm / 40 MOhm
-best accuracy: +/- (1%+2)
Frequency: 10Hz – 1MHz
– best accuracy: +/- (0.1%+3)
Duty cycle: 0.1% ~ 99.9%